Kalibrasi Temperature Gauge Menggunakan Oil Bath Temperature Calibrator CTB9500 Pada Industri Kilang

Authors

  • Dillon Diandri Politeknik Energi dan Mineral Akamigas
  • Farah Sukma Maulida Politeknik Energi dan Mineral Akamigas
  • Suka Handaja Politeknik Energi dan Mineral Akamigas

DOI:

https://doi.org/10.53026/2fv6hj95

Keywords:

kalibrasi, temperature gauge, kilang minyak, CTB9500, oil bath calibrator

Abstract

This study aims to analyze the calibration performance of temperature gauges using a WIKA CTB9500 oil bath calibrator—which features a stability of ±0.05°C and an accuracy of ±0.3°C—to improve temperature measurement accuracy in refinery processes. An experimental approach was employed, involving the calibration of 45 temperature gauge units from various brands with a measurement range of 0–60°C. The calibration process utilized a pre-cycle method at 0°C and 60°C, followed by testing in the sequence: 20°C → 30°C → 40°C → 30°C → 20°C. The calibration data were analyzed to determine error, hysteresis, and repeatability, with a pass criterion of a ±1°C tolerance. The results indicate that the CTB9500 was able to detect an average measurement error of 2.1°C in the "as-found" condition. Statistical analysis shows that 82% of the instruments met the ±1°C tolerance after calibration, while 18% were deemed unfit for further use (scrapped). Instruments in service for more than two years exhibited a drift of 0.38°C per month

References

[1] C. J. Delker, E. C. Auden, and O. M. Solomon, "Calculating Interval Uncertainties for Calibration Standards That Drift with Time," Journal of the IEST, vol. 63, no. 1, pp. 29-37, 2020.

[2] WIKA Instrumentation, "Calibration Bath or Dry-Well Calibrator? Advantages and Disadvantages," WIKA Blog, Sep. 2024. [Online]. Available: https://blog.wika.com/en/knowhow/calibration-bath-or-dry-well-calibrator-which-is-the-right-choice/

[3] A. Kumar, S. Patel, and R. Sharma, "Impact of Modern Calibration Equipment on Plant Reliability: A Case Study in Petrochemical Industry," J. Process Control, vol. 118, pp. 45-58, Sep. 2023.

[4] Thermo Electric Instrumentation, "Why Regular Calibration Is Critical for Reliable Temperature Measurement," Technical Article, Jan. 2025. [Online]. Available: https://www.thermo-electric.nl/why-regular-calibration-is-critical-for-reliable-temperature-measurement/

[5] Fluke Calibration, "Why Did My Temperature Sensor Fail Calibration?" Technical Papers and Articles, May 2024. [Online]. Available: https://us.flukecal.com/literature/articles-and-education/temperature-calibration/papers-articles/why-did-my-temperature-sen

[6] M. Singh, P. Johnson, and K. Williams, "Risk-Based Optimization of Calibration Intervals for Process Instruments," ISA Trans., vol. 142, pp. 234-248, Jan. 2024.

[7] Beamex, "How to Calibrate Temperature Sensors," Technical Blog, Sep. 2024. [Online]. Available: https://blog.beamex.com/how-to-calibrate-temperature-sensors

[8] Fluke Corporation, "Establishing Correct Calibration Intervals," Technical Literature, Jul. 2024. [Online]. Available: https://www.fluke.com/en-us/learn/blog/temperature-calibration/establishing-correct-calibration-intervals

[9] Mensor Corporation, "Deciding Between Calibration Baths and Dry Well Calibrators," Technical Blog, Apr. 2025. [Online]. Available: https://blog.mensor.com/blog/deciding-between-dry-wells-and-calibration-baths

[10] ISOBudgets, "How to Calculate Drift Uncertainty," Calibration Guidelines, Aug. 2025. [Online]. Available: https://www.isobudgets.com/calculate-drift/

Downloads

Published

2026-06-20

How to Cite

Kalibrasi Temperature Gauge Menggunakan Oil Bath Temperature Calibrator CTB9500 Pada Industri Kilang. (2026). Journal of Instrumentation and Hardware, 4(1), 13-21. https://doi.org/10.53026/2fv6hj95

Similar Articles

1-10 of 12

You may also start an advanced similarity search for this article.

Most read articles by the same author(s)